Part:BBa_K873002:Experience
This experience page is provided so that any user may enter their experience using this part.
Please enter
how you used this part and how it worked out.
Applications of BBa_K873002
User Reviews
UNIQ751c43c47e7619f1-partinfo-00000000-QINU
|};
Characterization
WashU_StLouis |
[http://2016.igem.org/Team:WashU_StLouis The WashU StLouis iGEM team] measured the strength of this HSP by attaching it to a GFP and measuring florescence at various temperatures. Protocol:
The plate in the plate reader was measured every fifteen minutes until the growth of the cells leveled out. This is observed when the absorbance of the cells stayed the same for multiple cycles. The fluorescence for the 30C and 42C plates were normalized and plotted over time. As seen in Figure 1 below, there was a spike in fluorescence shortly after the HSP was activated on the 42C plate. On the other hand, the HSP on the 30C plate was never activated. For both, fluorescence decreased greatly over time. At this time, we do not know if this is due to the HSP or the GFP device. The decrease in fluorescence over time may be due to photobleaching as well. Figure 2 below shows the measured fluorescence for each temperature before and after the experiment. Labels are given for Test 1 and Test 2 in order to differentiate between the two days the experiment was run. As expected, there is little difference between the temperatures before the experiment was run, but even after the HSP was heat activated, there was no discernible difference between the fluorescence of the various temperatures. In conclusion, we saw that this HSP biobrick was not activated by high temperatures. In order to reaffirm these results, [http://2016.igem.org/Team:Cardiff_Wales the Cardiff University iGEM team] repeated this protocol and got very similar results, as seen in Figure 3 and Figure 4.
|