Difference between revisions of "Part:BBa K2944003"
Line 8: | Line 8: | ||
===Results=== | ===Results=== | ||
<b>Characterization by Mass Spectrometry.</b><br> | <b>Characterization by Mass Spectrometry.</b><br> | ||
− | [[File:T--Concordia-Montreal--SilverResults1.png|400px|thumb|left|Figure 1 Mass Spectrometry]] | + | [[File:T--Concordia-Montreal--SilverResults1.png|400px|thumb|left|Figure 1 Mass Spectrometry Results. Image built using Prism 6. ]] |
<br><br> | <br><br> | ||
Line 15: | Line 15: | ||
<p> Activity of Glucose Oxidase was measured reacting the peroxide produce by the catalysis of glucose into gluconate with ammonium molybdate. The product of this reaction produces a yellow color that absorbs light at 405nm. The change in absorbance over time can be measured to determine if glucose oxidase is indeed present in the system. </p> | <p> Activity of Glucose Oxidase was measured reacting the peroxide produce by the catalysis of glucose into gluconate with ammonium molybdate. The product of this reaction produces a yellow color that absorbs light at 405nm. The change in absorbance over time can be measured to determine if glucose oxidase is indeed present in the system. </p> | ||
− | [[File:T--CONCORDIA-MONTREAL--Silver2.png|400px|thumb|left| | + | [[File:T--CONCORDIA-MONTREAL--Silver2.png|400px|thumb|left|<b>Figure 2 </b>– Absorbance of Cell Supernatant Over Time in the Presence of 100mM Ammonium Molybdate.]] |
− | <b>Figure 2 </b>– Absorbance of Cell Supernatant Over Time in the Presence of 100mM Ammonium Molybdate.<br><br> | + | <br><br> |
The sample was subjected to a solution of 10mM D-glucose and 100mM ammonium molybdate. The absorbance was measured periodically over a time frame of three hours.<br><br> | The sample was subjected to a solution of 10mM D-glucose and 100mM ammonium molybdate. The absorbance was measured periodically over a time frame of three hours.<br><br> | ||
− | [[File:T--Concordia-Montreal--Silver3.png|400px|thumb|left| | + | [[File:T--Concordia-Montreal--Silver3.png|400px|thumb|left|<b>Figure 3</b> Calibration Curve of the Absorbance at 405nm of Ammonium Molybdate]] |
− | <b>Figure 3</b> Calibration Curve of the Absorbance at 405nm of Ammonium Molybdate<br><br><br> | + | <br><br><br> |
− | |||
− | [[File:T--Concordia-Montreal--Silver4.png|400px|thumb|left|Table 1]] | + | [[File:T--Concordia-Montreal--Silver4.png|400px|thumb|left|<b> Table 1-</b> Limit of Detection and Quantification of the Calibration Curve]] |
<br><br> | <br><br> | ||
Line 79: | Line 78: | ||
In HPLC-MS, samples are purified using HPLC and are then analyzed using mass spectroscopy to determine the compound. This system can not only identify samples with high accuracy, but also quantitate them based on the area of the peak producing the signal of interest.<br><br> | In HPLC-MS, samples are purified using HPLC and are then analyzed using mass spectroscopy to determine the compound. This system can not only identify samples with high accuracy, but also quantitate them based on the area of the peak producing the signal of interest.<br><br> | ||
− | + | ||
− | [[File:T--Concordia-Montreal--HPLC1.png|400px|thumb|left|]] | + | [[File:T--Concordia-Montreal--HPLC1.png|400px|thumb|left|<b>Table 1</b>Calibration Data]] |
<br> | <br> | ||
− | [[File:T--Concordia-Montreal--SilverGluconolactone.png|400px|thumb|left| | + | [[File:T--Concordia-Montreal--SilverGluconolactone.png|400px|thumb|left|<b>Figure 1-</b>Figure 4- Standard Curves of Gluconolactone and Gluconate]] |
− | <b>Figure 1-</b>Figure 4- Standard Curves of Gluconolactone and Gluconate<br> | + | <br><br> |
− | + | ||
− | + | ||
− | [[File:T--COncordia-Montreal--table3HPLC.png|400px|thumb|left|]] | + | [[File:T--COncordia-Montreal--table3HPLC.png|400px|thumb|left|<b>Table 2-</b> Sample Signals and Determined Concentration]] |
<br><br>Both samples show the production of gluconate although these values are low, they are still able to be quantified within the range of the calibration curve. The most likely reason for low production is a population of cells in the sample.<br> | <br><br>Both samples show the production of gluconate although these values are low, they are still able to be quantified within the range of the calibration curve. The most likely reason for low production is a population of cells in the sample.<br> | ||
<br> | <br> | ||
<b>Results</b> | <b>Results</b> | ||
− | [[File:T--Concordia-Montreal--SilverResults1.png|400px|thumb|left|Figure 1 Mass Spectrometry Results | + | [[File:T--Concordia-Montreal--SilverResults1.png|400px|thumb|left|Figure 1 Mass Spectrometry Results. Image built using Prism 6.]] |
− | Image built using Prism 6. | + | |
There are two superimposed traces: gluconolactone standard and yeast supernatant mixed 50/50 with 40 g/L glucose. | There are two superimposed traces: gluconolactone standard and yeast supernatant mixed 50/50 with 40 g/L glucose. |
Revision as of 00:17, 22 October 2019
pTDH3-GOx-tPGK1
This coding sequences encodes a constitutive yeast promoter regulating expression of the glucose oxidase enzyme from Aspergillus niger. The gene sequence has been optimized for Saccharomyces cerevisiae. Glucose oxidase catalyzes the oxidation of glucose to D-glucono-1,5-lactone and hydrogen peroxide.
Results
Characterization by Mass Spectrometry.
In HPLC-MS, samples are purified using HPLC and are then analyzed using mass spectroscopy to determine the compound. This system can not only identify samples with high accuracy, but also quantitate them based on the area of the peak producing the signal of interest.
Preparation of Standard Curve
Activity of Glucose Oxidase was measured reacting the peroxide produce by the catalysis of glucose into gluconate with ammonium molybdate. The product of this reaction produces a yellow color that absorbs light at 405nm. The change in absorbance over time can be measured to determine if glucose oxidase is indeed present in the system.
The sample was subjected to a solution of 10mM D-glucose and 100mM ammonium molybdate. The absorbance was measured periodically over a time frame of three hours.
The change in absorbance indicates that glucose oxidase is present in the system and can be detected. Furthermore, the similarity in the rate of change in absorbance further suggests the presence of enzymatic activity. However, the level of peroxide detected is too low to be used to accurately quantified the concentration of peroxide in the system and the amount of glucose oxidase in the sample. This is likely due to the low cell concentration in the sample Which was approximately 1.74*107 cells/ml.
Experimental
Experimental Conditions
Sample prep
Standards
Authentic gluconolactone and gluconate standards were purchased from Sigma-Aldrich
Samples
Samples diluted 1:5 in cold acetonitrile and centrifuged at 21,000 RCF for 2 minutes
5 μL was injected into HPLC
HPLC Conditions
Machine:
1290 Infinity II LC system (Agilent Technologies)
Column:
Zorbax Eclipse Plus C18 column (50 x 2.1 mm, 1.8 uM; Agilent Technologies)
Column Temperature:
30°C
Solvents:
A: 0.1% formic acid in water
B: 0.1% formic acid in acetonitrile
Gradient:
2%B for 1 min, 85%B for 2 min, followed by re-equilibration for 2 min
MS Conditions
Machine:
Agilent 6545 quadrupole time-of-flight MS (QTOF-MS; Agilent Technologies)
Gas Settings:
Sheath gas flow rate: 10 L / min
Sheath gas temperature: 350 C
Drying gas flow rate: 12 L / min
Drying gas temperature: 325 C
Nebulizing gas: 55 psig
Ionization
Negative mode
Sample Analysis
Programs:
Agilent MassHunter Qualitative Analysis software
Agilent MassHunter Quantitative Analysis software
Compound confirmation
Exact mass of gluconolactone, [M-H]- : 177.0399
Exact mass of gluconate, [M-H]- : 195.0505
Samples were confirmed by comparison of retention time and exact mass to authentic standard
Exact mass was accurate to < 10 ppm
In HPLC-MS, samples are purified using HPLC and are then analyzed using mass spectroscopy to determine the compound. This system can not only identify samples with high accuracy, but also quantitate them based on the area of the peak producing the signal of interest.
Both samples show the production of gluconate although these values are low, they are still able to be quantified within the range of the calibration curve. The most likely reason for low production is a population of cells in the sample.
Results
There are two superimposed traces: gluconolactone standard and yeast supernatant mixed 50/50 with 40 g/L glucose. The traces are of extracted ion 195.0505 (exact [M-H]- of gluconate. The gluconolactone standard is predominately gluconate upon resuspension in water.
The mass spectrum in the inset is the total ion count of the standard at 0.5 mins between the indicated m/z range. The two masses which are not the exact mass of gluconate are also part of the standard. Their identities are confirmed using the online database mzCloud. Annotated in the figure is we believe these masses correspond to based on the papers referenced below.
- Characterization done by Matthew Tiranardi.
Thankyou Lauren Narcross for her technical assistance!
References:
Taylor, V. F., March, R. E., Longerich, H. P., & Stadey, C. J. (2005). A mass spectrometric study of glucose, sucrose, and fructose using an inductively coupled plasma and electrospray ionization. International Journal of Mass Spectrometry, 243(1), 71–84. doi: 10.1016/j.ijms.2005.01.001
Zhang, Z., Gibson, P., Clark, S. B., Tian, G., Zanonato, P. L., & Rao, L. (2007). Lactonization and Protonation of Gluconic Acid: A Thermodynamic and Kinetic Study by Potentiometry, NMR and ESI-MS. Journal of Solution Chemistry, 36(10), 1187–1200. doi: 10.1007/s10953-007-9182-x
Sequence and Features
- 10COMPATIBLE WITH RFC[10]
- 12COMPATIBLE WITH RFC[12]
- 21INCOMPATIBLE WITH RFC[21]Illegal BamHI site found at 717
Illegal BamHI site found at 1027 - 23COMPATIBLE WITH RFC[23]
- 25COMPATIBLE WITH RFC[25]
- 1000INCOMPATIBLE WITH RFC[1000]Illegal BsaI.rc site found at 1051
Illegal BsaI.rc site found at 2269